HI8614 pH Transmitter with 4-20 mA Galvanically Isolated Output
RM 3,133.00
1.500 KG
Select Option
The HI8614N is a water-resistant pH transmitter is designed to be used with a standard high impedance pH probe with BNC connector. The signal is then processed by a special highimpedance amplifier, which transmits an output current directly proportional to the input signal but independent of changes in load or cable capacitance.
- Temperature compensation ensures fast, accurately compensated readings even during sudden temperature fluctuations.
- Water resistant design for use in process control.
- Optional backlit, LCD display for easy viewing.
Automatic Temperature Compensation (ATC)
- Temperature compensation is performed by the transmitter’s ATC (Automatic Temperature Compensation) circuitry when measurements are taken with a temperature probe attached; if ATC is not required, it is also possible to substitute a fixed resistor for the temperature probe.
BNC Connector
- The transmitters use a universal BNC socket for quick and secure connection to any electrode with a BNC connector.
Easy Calibration
- Calibration is performed by the adjustment of two independent trimmers — slope and offset.
Isolated Input
- The input is isolated from the current loop to eliminate problems related to ground loop, low insulation cables, multiple electrode connections, and a common mode voltage of up to 100V ensures true differential readings
Flexible for Your Needs
- The transmitter can be connected to any Hanna controller that accepts analog input.
Version with LCD
- HI8614”L” versions allow easy verification and monitoring of measured values and is easier to calibrate and maintain.
pH
pH Range | 0.00 to 14.00 pH; 4–20 mA |
---|---|
pH Resolution | (for “L” models) 0.01 pH; 0.01 mA |
pH Accuracy | ±0.02 pH; ±0.02 mA |
pH Calibration | offset: ±2 pH; ±2.2 mA; slope: 86 to 116%; ±0.5 mA |
pH Temperature Compensation | fixed or automatic from 0 to 100°C (32 to 212°F) with HI76608 probe |
pH Input impedence | 1012 ohm |
Process Control Meter
Recorder Output | 4-20 mA (isolated) |
---|---|
Load | max 500 Ohm |
Display | LCD only for HI8614LN |
Power Supply | HI8614N: 18-30 VDC; HI8614LN: 20-36 VDC |
Environment | 0 to 50°C (32 to 122°F); RH max 95% non-condensing |
IP Rating | IP65 |
Dimensions | 165 x 110 x 71 mm (6.5 x 4.3 x 2.8’’) |
Weight | 1 kg (2.2 lb.) |
Ordering Information | HI8614N and HI8614LN (with display) is supplied with instructions. |