HI8615 ORP Transmitter with 4-20 mA Galvanically Isolated Output
HI8615 ORP Transmitter with 4-20 mA Galvanically Isolated Output Controllers, Analyzers & Transmitters Process Control Selangor, Malaysia, Penang, Sabah, Kuala Lumpur (KL), Petaling Jaya (PJ), Sungai Nibong, Kota Kinabalu Supplier, Suppliers, Supply, Supplies | Hanna Instruments (M) Sdn Bhd
RM 3,133.00
1.500 KG
1 in stock
Select Option
Download Files
manhi_8614_8615.pdf (574.84 KB)
manhi_8614_8615.pdf (574.84 KB)
The HI 8615 has been designed for transmitting ORP measurements from remote locations in industrial environments. It features two controls (one for 4 mA and one for 20 mA) to compensate for electronic drift and ambient temperature.

  • Designed for transmitting ORP measurements from remote locations in industrial environments.
  • Water resistant design for use in process control.
  • Ideal for chromium hexavalent reductions, cyanide oxidation reactions and water sanitation.

Robust Design

  • IP65 rating and a rugged polypropylene casing provide optimum protection even in harsh environments, making this instrument ideal for chromium hexavalent reductions, cyanide oxidation reactions and water sanitation.

BNC Connector

  • The transmitters use a universal BNC socket for quick and secure connection to any electrode with a BNC connector.

Easy Calibration

  • Calibration is performed by the adjustment of two independent trimmers — slope and offset.

Isolated Input

  • The input is isolated from the current loop to eliminate problems related to ground loop, low insulation cables, multiple electrode connections, and a common mode voltage of up to 100V ensures true differential readings

Flexible for Your Needs

  • The transmitter can be connected to any Hanna controller that accepts analog input.

Version with LCD

  • HI8615 ”L” versions allow easy verification and monitoring of measured values and is easier to calibrate and maintain.

mV

mV Range 0 to ±1000 mV; 4–20 mA
mV Resolution (for “L” models) 1 mV; 0.01 mA
mV Accuracy ±5 mV; ±0.02 mA
mV (ORP) Calibration offset: ±0.8 mA/±100mV; slope: 90 to 110%; ±0.8 mA

Process Control Meter

Recorder Output 4-20 mA (isolated)
Load max 500 Ohm
Display LCD only for HI8615LN
Power Supply HI8615N: 18-30 VDC; HI8615LN: 20-36 VDC
Environment 0 to 50°C (32 to 122°F); RH max 95% non-condensing
IP Rating IP65
Dimensions 165 x 110 x 71 mm (6.5 x 4.3 x 2.8’’)
Weight 1 kg (2.2 lb.)
Ordering Information HI8615N and HI8615LN (with display) is supplied with instructions.

Please leave your enquiry here, we will reply as soon as possible.