HI4108 Copper Combination Ion Selective Electrode (ISE)
HI4108 Copper Combination Ion Selective Electrode (ISE) ISE Electrodes & Probes Selangor, Malaysia, Penang, Sabah, Kuala Lumpur (KL), Petaling Jaya (PJ), Sungai Nibong, Kota Kinabalu Supplier, Suppliers, Supply, Supplies | Hanna Instruments (M) Sdn Bhd
RM 5,835.00
500 g
1 in stock
Download Files
manhi_4108.pdf (194.33 KB)
The HI4108 is a solid state, combination ion selective electrode (ISE) for the determination of cupric ions (Cu2+) in solution. The mixed cupric sulfide/silver sulfide sensing pellet produces a potential change due to the cupric ion exchange between the inorganic pellet membrane and the sample. The internal sensing elements are housed within a durable polyetherimide (PEI) body.

 

The HI4108 is ideal for a variety of applications in plating, wastewater, and industrial processes.


  • Solid state sensor
  • Skirted cone geometry
  • Detection from 0.065 to 6,355 mg/L Cu2+

Details

The HI4108 cupric ion selective electrode is a complete potentiometric cell that contains a silver/silver chloride (Ag/AgCl) reference housed within a PEI plastic body with a potassium nitrate electrolyte. The internal elements are isolated from the sample by a compressed cupric sulfide/silver sulfide sensing membrane.

Hanna’s solid state combination ISEs are designed with a cone style junction. The skirted cone geometry of the sensor forms the liquid junction with the test solution at the tip of the junction cone, producing a highly stable reference potential and consistent flow rates of electrolyte into the solution. A simple press of the top portion of the ISE makes it easy to empty the electrolyte fill solution and replace when necessary.

For the HI4108 to measure cupric ions accurately, it is important that the ionic strength of the standards and sample be adjusted to a high and constant value with the HI4000-00 ionic strength adjuster (ISA) solution. The ISA ensures that the activity coefficient is constant so the concentration of free cupric ions can be measured. Measuring all solutions with a constant ionic strength reduces the margin of error between measurements. The HI4108 cupric solid state ISE can measure from 1µM (0.065 mg/L) to 0.1M (6,355 mg/L).

 

Conversion Table for Cu2+

 

Multiply

moles/L (M) to ppm (mg/L)

6.354 x 104

ppm (mg/L) to M (moles/L)

1.574 x 10-4

 
 

Features at-a-glance

Fixed Sensing Modules - Combination solid state ISEs are electrodes that detect free ions in a solution. The highly insoluble, compressed sensing pellet is attached to the electrode by a fixed sensing module for ease of use.

Durable PEI Body - The probe body is composed of durable polyetherimide (PEI) plastic.

BNC Connection - The HI4108 has a universal BNC connector for easy connection to any benchtop meter with a female BNC probe input.


Theory of Operation

A solid state sensor electrode develops a voltage due to the ion exchange that occurs between the solution and the inorganic membrane. An equilibrium occurs due to the highly limited solubility of the membrane material in the sample. When the ionic strength of the solution is fixed by the addition of ISA, the voltage is proportional to the concentration of free ions in solution.

Specifications

Type combination
Measurement Range Cupric (Cu2+) 0.1M to 1•10-6M, 6,355 to 0.065 mg/L (ppm)
Optimum pH Range pH 3 to 7
Temperature Range 0 to 80 °C
Approximate Slope +27 mV
Diameter 12 mm
Overall Length 120 mm
Body Material PEI
Cable coaxial; 1 m (3.3’)
Connection BNC
   

 

 