HI3222 is a professional dual-input pH/ORP/ISE/Temperature Bench Meter with a graphic LCD designed to provide high accuracy and ease of use both in the laboratory as well as in harsh industrial conditions. This instrument is a dual-input model measuring pH or ORP and temperature on channel 1 and ISE or ORP and temperature on channel 2. A Relative mV feature is also available.

Main Features:
• Dual-input channel model
• Up to 5 point pH calibration with 7 standard buffers and 5 custom buffers to choose from
• Messages on the graphic LCD for an easy and accurate calibration
• Diagnostic features to alert the user when the electrode needs cleaning
• Three types of logging modes (for pH, ISE and mV)
- Stability logging
- Interval logging
- Log-on-demand
• Relative mV measurements
• Log on demand up to 400 samples
• Log interval with log on stability feature up to 600 records
• Auto Hold feature, to freeze first stable reading on the LCD
• GLP feature, to view last calibration data for pH, Rel mV or ISE
• PC interface

More Information
SKU HI3222
Product Name Dual-Input pH/ORP/ISE/Temperature Bench Meter with Calibration Check - HI3222
Quote Required Yes
pH Range -2.0 to 20.0 pH / -2.00 to 20.00 pH / -2.000 to 20.000 pH
pH Resolution 0.1 pH / 0.01 pH / 0.001 pH
pH Accuracy ±0.1 pH / ±0.01 pH / ±0.001 pH
pH Calibration Up to 5 points / 7 standard buffers (1.68, 4.01, 6.86, 7.01, 9.18, 10.01, 12.45) / 5 custom buffers
ORP Range ±2000.0 mV
ORP Resolution 0.1 mV
ORP Accuracy ±0.2 mV
ISE Range 1.00xE-7 to 9.99xE10 conc.
ISE Resolution 3 digits, 0.01, 0.1, 1, 10 conc.
ISE Accuracy ±0.5% of reading (monovalent ions) / ±1% of reading (divalent ions)
ISE Calibration Up to 5 points / 6 standards (0.1, 1, 10, 100, 1000, 10000 ppm)
Temperature Range -20 to 120 °C (-4.0 to 248.0 °F)
Temperature Resolution 0.1 °C (0.1 °F)
Temperature Accuracy ±0.2 °C (±0.4 °F) (excluding probe error)
Rel mV offset range ±2000 mV
Slope Calibration From 80 to 110%
Temperature compensation Manual / Automatic
Electrode Hanna BNC Connection pH / ORP / ISE electrodes
Temperature Probe RCA connection / Recommended option
LOG on demand 400 samples
Lot logging 5, 10, 30 seconds / 1, 2, 5, 10, 15, 30, 60, 120, 180 minutes / AutoEnd (maximum 600 samples)
Environment 0 to 50 °C (32 to 122 °F) / max. RH 55% non-condensing
Power supply 12 Vdc power adapter
PC interface opto-isolated USB
Dimensions 235 x 207 x 110 mm (9.2 x 8.14 x 4.33 in.)
Weight (Meter only) 1.8 Kg (4.1lbs)