- Digital input for pH and ORP probes
- Up to 5 relays for control and cleaning sequence
- Up to 4 analog output for control and monitoring
HI510 is an advanced universal process controller that can be configured for many applications requiring monitoring and/or control of process parameters. It features a digital probe input that will automatically detect and update the controller with the parameter that it measures.
This controller can be configured with up to 5 control relays and 4 analog outputs. The control relays can be programmed for on/off, Proportional, or PID (Proportional Integral Derivative) types of control or used for programming cleaning sequences triggered by a digital input or based on a schedule. The scalable analog outputs can be used for remote devices including Chart recorders, PLC, or SCADA. The analog outputs can be used for control devices including dosing pumps and valves that accept a 0/4-20 mA output.
HI510 features a RS485 digital output that allows up to 32 devices to be connected to a single line that can be viewed and programmed remotely using the HI92500 software. The HI92500software provides for graphic user interface that allows for settings to be changed and saved for transfer to another HI510 controller.
The HI510 NEMA 4X enclosure uses sealed cable glands to maintain waterproof integrity. The versatile mounting kits allow for mounting the HI510 to a wall, pipe and panel.
More Information
SKU | HI510 |
---|---|
Product Name | Universal Process Controller - HI510 |
Quote Required | No |
Dosing Control | on/off, proportional and PID |
Dosing Terminals |
Up to 5 relays (independently configurable for process variables, Hold & Cleaning functions) Electromechanical relay SPDT and SPST contact outputs 5A – 250 Vac; 5A – 30 Vdc (resistive load) Fuse protected |
Alarm Relay Set Point |
Up to 5 relays (independently configurable for process variables, Hold & Cleaning functions) Electromechanical relay SPDT and SPST contact outputs 5A – 250 Vac; 5A – 30 Vdc (resistive load) Fuse protected |
Over Dosing Control | yes |
System Error | yes |
Analog Ouput Signal |
2 or 4 independent, galvanically isolated outputs 0 – 22 mA configurable as |
Analog Ouput Accuracy | ±2% full scale |
Digital Input Signal |
independent, galvanically isolated inputs (configurable for Hold & Cleaning functions) On state |
Digital Output Signal | RS485 |
Backlight | yes |
Power Consumption | 15VA |
Enclosure | *Single case ½ DIN, type NEMA4X, IP65 ingress protection |
Installation Category | II |
Panel Cutout | 1/2 DIN (brackets ordered seperately) |
Input Channels | one digital input with RS485 interface |
pH Electrode | HI10X6 series – pH and Temperature (ordered seperately) |
ORP Electrode | HI20X4 series – ORP and Temperature (ordered seperately) |
Logging Type |
Interval log, up to 100 files, maximum 8600 records on each stored file. When the maximum limit of 100 stored files is reached, the most recent file will automatically erase the oldest one. Event log, maximum 100 records. When the maximum limit i |
Connectivity |
RS485 serial port - Remote monitoring and control USB-C port - Retrieve log files and firmware upgrading |
GLP | yes |
Calibration Timeout Reminder | yes |
Analog Output | up to 4 based on configuration ordered |
Display | Graphic LCD, 128 x 64 pixel B/W with backlight |
Power Supply | 100 – 240 Vac ±10%; 50/60 Hz; 15VA; fuse protected (2A, 250V slow blow fuse) |
Environment | -20 to 50 ºC (-4 to 122 ºF); maximum 100% RH non-condensing |
IP Rating | IP65 |
Dimensions | Width 144.0 mm (5.7”) x Height 144.0 mm (5.7”) x Depth 151.3 mm (6.0”) |
Weight | Approximately 1.6 kg (3.5 lb.) |
Notes | *For a water tight seal, tighten the front four screws on the front casing to 13.3 lbf·in (1.5 N·m, maximum 2.0 N·m), of torque. |