HI3221 pH/ORP/ISE/Temperature Bench Meter with Calibration Check
RM 2,408.00
SKU: HI3221
1.800 KG
23 cm (L) x 10 cm (W)
HI3221 is a professional pH/ORP/ISE/Temperature bench meter with graphic LCD designed to provide high accuracy and ease of both in the laboratory as well as in harsh industrial conditions. This instrument is a single input model measuring pH, ORP, or ISE and Temperature utilizing a single channel.
Main Features:
Main Features:
- Single input channel
- Up to 5 point pH calibration with 7 standard buffers and 5 custom buffers to choose from
- Messages on the graphic LCD for easy and accurate calibration
- Diagnostic features to alert the user when the electrode needs cleaning
-
Three types of logging modes:
- Stability logging
- Interval logging
- Log-on-demand
- Relative mV measurements
- Log on demand up to 300 samples
- Log interval with log on stability feature up to 600 records
- Auto Hold feature, to freeze the first stable reading on the LCD
- GLP feature, to view last calibration data for pH, Rel mV, or ISE
- PC interface
SKU | HI3221-02 |
---|---|
Product Name | pH/ORP/ISE/Temperature Bench Meter with Calibration Check - HI3221 |
Quote Required | Yes |
pH Range | -2.0 to 20.0 pH / -2.00 to 20.00 pH / -2.000 to 20.000 pH |
pH Resolution | 0.1 pH / 0,01 pH / 0.001 pH |
pH Accuracy | ±0.1 pH / ±0.01 pH / ±0.002 pH |
pH Calibration | Up to 5 points / 7 Standard buffers (1.68, 4.01, 6.86, 7.01, 9.18, 10.01, 12.45) / 5 Customer buffers |
ORP Range | ±2000.0 mV |
ORP Resolution | 0.1 mV |
ORP Accuracy | ±0.2 mV |
ISE Range | 1.00xE-3 to 1.00xE5 conc. |
ISE Resolution | 3 digits, 0.01, 0.1, 1, 10 conc. |
ISE Accuracy | ±0.5% of reading (monovalent ions) / ±1% of reading (divalent ions) |
ISE Calibration | Up to 2 points / 6 standards (0.1, 1, 10, 100, 1,000, 10,000 ppm) |
Temperature Range | -20 to 120.0 °C (-4.0 to 248.0 °F) |
Temperature Resolution | 0.1 °C (0.1 °F) |
Temperature Accuracy | ±0.2 °C (±0.4 ° F) (excluding probe error) |
Rel mV offset range | ± 2000 mV |
Slope Calibration | From 80 to 110% |
Temperature Compensation | Manual / Automatic |
Electrode | Hanna BNC connection pH / ORP / ISE electrodes |
Temperature Probe | RCA connection / Recommended option |
LOG on demand | 300 samples |
Lot logging | 5, 10, 30 seconds / 1, 2 ,5, 10, 15, 30, 60, 120, 180 minutes, AutoEnd (maximum 600 samples) |
Environment | 0 to 50 °C (32 to 122 ° F) max. RH 55% non-condensing |
Power Supply | 12 Vdc power adapter |
PC Interface | opto-isolated USB |
Dimensions | 235 x 207 x 110 mm (9.2 x 8.14 x 4.33 in.) |
Weight (Meter Only) | 1.8 kg (4.1 lbs) |
Ordering Information | Supplied with 12 Vdc Power Adapter, Quick Reference guide with instrument quality certificate |